Measurement of retentive forces of three different attachments in shape |
ZUO En - jun1, DUAN Wei-hong2, GAO Shi-jun2, LIU Ming-da2
|
1.Stomatology College, Dalian Medical University,Dalian 116044,China;2.Stomatology College,Jiamusi University, Jiamusi 154002,China
|
Abstract: |
[Objective]To measure retentive forces of three different attachments in shape.[Methods]The attachments were divided into five groups.The retentive force specimens of five groups were made respectively. Then each one was drawn vertically ten times on mechanical test-bed. The data were analyzed statistically.[Results]Retentive forces of five grpups of attachments:A:(10.26±0.34)N, B: (8.22±0.41)N; C:(7.96±0.29) N; D:(10.47±0.61)N; E:(10.52±0.58)N. There was significant difference between A and B or C ,B and D,C and E (P<0.05), There was no significant difference between A and D or E , D and E(P>0.05).[Conclusion]Retentive forces of three different attachments all meet clinical requirement. Stud-snap attachment with white plastic cap can give the highest retentive force. |
Key words: attachment retentive force hardness |