摘要: |
[目的] 比较两种不同的方法测定红细胞沉降率(ESR)的结果,探讨自动血沉仪的临床应用。[方法]评价Micro-test1血沉仪的精密度,比较自动血沉仪法和传统魏氏法对413例临床血液标本同步进行测定ESR的结果并进行分析。[结果]Micro-test1血沉仪具有较好的精密度,中值样本和高值样本的变异系数(CV)<5.0%。两种方法测定ESR都升高有199例,都正常214例,两法符合率为100%,但分析升高样本时,发现传统的魏氏法超过90 mm/h的有15例,而自动血沉仪法结果均在90 mm/h以下。[结论]Micro-test1自动血沉仪法和魏氏法测定ESR正常和升高的结果符合率为100%,但在ESR超过90 mm/h时,两者结果有差异,造成此差异的原因可能与本仪器有关。 |
关键词: 红细胞沉降率 魏氏法 Micro-test1血沉仪 |
DOI:10.11724/jdmu.2008.03.26 |
分类号:R446 |
基金项目: |
|
Evaluation of comparison between Wei Shi methods and Micro-test1 analyzer in detecting ESR |
GONG Ai - hua, SUN Li - juan, DONG Feng, HE Xiao - lin
|
Department of Clinical Laboratory, the Second Affiliated Hospital of Dalian Medical University, Dalian 116027, China
|
Abstract: |
[Objective]To compare ESR results of two different methods and to study the clinical application of automatic ESR analyzer.[Methods]Precision of micro-test1 analyzer was detected .413 blood samples'ESR were detected symutaneously with analyzer and manual methods .The results were analyzed.[Results]Micro-test1 analyzer had qualified precision. The coefficient of variation (CV) of moderate and high test results were all less than 5.0%. 199 cases 'ESR increased and 214 cases were normal with above two methods. Their according rates were 100%. Among samples whose ESR increased, 15 cases 'ESR excelled 90mm/h with classical Wei Shi methods and none excelled 90 mm/h with analyzer.[Conclusions]according rates of micro-test1 analyzer and Wei Shi methods are 100%, But there is difference between these two methods when ESR is above 90 mm/h,the reason of it maybe correlated with analyzer. |
Key words: ESR Wei Shi methods Micro-test1 ESR analyzer |