摘要: |
[目的] 用即经济又可靠的方法对本室CA系列血凝仪进行质量控制。[方法] 收集凝血像正常的住院患者的混和血浆,取出3 mL,其余分装,-70 ℃冰冻保存,按方法学评价的内容进行批内、日间精密度,稳定性观察,并定期用进口定值质控品LEVER1观察其准确性。[结果]批内精密度凝血酶原时间(PT):1.3% 、部分凝血活酶时间 (APTT):1.9% 、 纤维蛋白原(Fbg):3.1%.。日间精密度 CV% PT<4% 、APTT<4% 、Fbg<7% ,符合凝血项目质控要求。定期测定定值质控品值均在定值范围,准确性良好。[结论] 自制正常范围质控品有良好精密度、稳定性、准确性,可代替进口正常范围质控品用于CA血凝仪室内质控。 |
关键词: 自制正常范围质控品 室内质控 CA-7000血凝仪 |
DOI:10.11724/jdmu.2010.01.26 |
分类号:R446 |
基金项目: |
|
Self-made normal range control as sysmex CA-coagulant-meter indoor quality control |
JIN Yan, LI Yan-lian, ZHANG Feng-hua, QIN Guei-e
|
Department of Clinical Laboratory, the First Affiliated Hospital of Dalian Medical University, Dalian 116011, China
|
Abstract: |
[Objective] To make quality control on sysmex CA-coagulant-meter with an economic and dependable method. [Methods] The mixed plasma was collected from hospitalized patients with normal coagulation time. After taking 3 ml plasma out, the rest were kept at -70℃. Intra-run CV%, inter-day CV% and stability were observed according to the contents of methodology evaluation. The accuracy was monitored by the imported packaged control. [Results] Intra-run CV% showed that plasma prothrombin time (PT), activated partial thromboplasting time (APTT) and fibrinogen (Fbg) were 1.3%, 4.6% and 3.1% respectively. Inter-day CV% indicated that PT, APTT and Fbg were less than 4%, 4% and 7% respectively. These indexes met the requirement of coagulation item control. All of the periodically measured values were within the assigned value ranges with high accuracy. [Conclusion] Self-made normal range control has high precision, accuracy and stability. It is acceptable that self-made normal range control may replace imported package control as sysmex CA-coagulant-meter indoor quality control. |
Key words: self-made normal range quality control indoor quality control sysmex CA-coagulant-meter |